UC8_R_21 https://repo.valu3s.eu/use-cases/neuromuscular-transmission-measurements/requirements/uc8_r_21 https://repo.valu3s.eu/@@site-logo/logo_valu3s_green_transparent.png UC8_R_21 Requirement ID 221 Requirement Textual Description Model outputs are within expected ranges in time domain and in NMT domain (TOF values). Detailed specification: error in estimating the intubation time is <= 20%, error in estimating half-life time is <= 20%. Requirement Type Functional Requirement Test Cases Requirement Test Cases List UC8_TC_5 The Controller calculates the new relaxant dose value to inject and sends it to the Infusion pump and to the HiL Manager Tool. Controller generates random sequences of infusion (initial bolus, continuous infusion). Recovery time to half concentration of drug in plasma is in expected range (half-life), error is < 20%; Intubation time is in expected range, error is < 20% VALU3S Framework Evaluation Environment Type Type of environment where the method is applied. It can be more than one type of environment.(Dimension 1) Evaluation Type Type of evaluation performed with that method. If a method is hybrid, more than one type can be selected.(Dimension 2) Type of Component Under Evaluation Type of components that can be evaluated with that method. It can be more than one type.(Dimension 3) Evaluation Stage The evaluation stage where the method is used.(Dimension 5) Purpose of the Component Under Evaluation The type of purpose that can be evaluated by a method. It can be more than one type.(Dimension 6) Type of Requirement Under Evaluation The type of requirements that a method is able to evaluate. It can be more than one type.(Dimension 7) Evaluation Performance Indicator The type of evaluation criteria or KPI that a method is able to improve. It can be more than one type.(Dimension 8) Contents There are currently no items in this folder.